MDClassifyStart free classification

GUIDE

MDR or IVDR: which regulation applies to your product

Two regulations, adopted the same day, with different class letters, different rules and different consequences. The line between them is narrower than most people expect — and getting it wrong means every decision downstream was made under the wrong framework.

Written with Francesca Di Giuseppe, Senior QA/RA Specialist (LinkedIn) — content reviewed on 1 August 2026

acts on the patientexamines a specimen

The question that decides it

Does the product examine a specimen taken from the human body, outside the body?

If yes, it is an in vitro diagnostic and the IVDR applies. If no, and it is a medical device at all, the MDR applies. Almost every borderline case is an argument about that one sentence.

What counts as a specimen

Blood, tissue, urine, saliva, swabs — anything derived from the body and examined outside it. The examination must be intended to provide information about a physiological or pathological state, a congenital abnormality, a predisposition, the compatibility with a potential recipient, or the monitoring of therapeutic measures.

NOTE —

Specimen receptacles are expressly treated as in vitro diagnostics. A tube that only holds a sample, and does nothing else, is still governed by the IVDR — a detail that surprises manufacturers of consumables.

Acting on the patient versus acting on a sample

The clearest way to hold the distinction: an MDR device does something to or for the patient. An IVDR device tells you something about a sample that came from the patient. A glucose meter measuring a drop of blood is an in vitro diagnostic. A continuous glucose monitor with a sensor in the tissue is not — it never leaves the body, so nothing was examined in vitro.

Why the answer changes everything after it

MDR — 2017/745IVDR — 2017/746
ClassesI · IIa · IIb · IIIA · B · C · D
Classification rules22 rules in Annex VIII7 rules in Annex VIII of the IVDR
What the rules weighInvasiveness, duration of contact, whether the device is active, special categoriesWhat the test detects, and the consequence of a wrong result — for the patient and for public health
Highest classIII — highest individual riskD — highest individual and public health risk
Lowest classI — self-declared, with the Is/Im/Ir exceptionsA — self-declared, unless supplied sterile

The public health dimension has no MDR equivalent

This is the conceptual difference worth understanding rather than memorising. An MDR device is assessed on the harm it can do to the person using it. An IVD can be dangerous to people who never touch it: a test that fails to detect a transmissible agent in donated blood endangers a recipient, and potentially a population. The IVDR therefore weighs individual risk and public health risk together, which is why the highest class is reserved for tests on transmissible agents and life-threatening conditions with a high propagation risk.

The Notified Body consequence

Under the previous directive, the large majority of in vitro diagnostics were self-declared with no Notified Body. Under the IVDR only Class A devices are, and Class A is a narrow category — instruments, receptacles, buffers and washing solutions. The proportion of IVDs requiring a Notified Body inverted, which is the main reason IVDR transition has been harder than MDR transition.

NOTE —

The transitional provisions of both regulations have been amended several times since adoption. Any deadline you are working to should be checked against the current consolidated text rather than an article written when it was first published — including this one.

Where products sit awkwardly

Software follows the data

Software is where the line is thinnest. It is not the software that is examined, so ask what it works on. Software interpreting the output of an in vitro examination is governed by the IVDR. Software working on images, vital signs, symptoms or records is governed by the MDR. The endorsed guidance on qualification and classification of software under both regulations is MDCG 2019-11.

A platform doing both is not one product with one answer. Each function is qualified on its own, and a company can end up holding certificates under both regulations.

Companion diagnostics

A test that determines whether a patient is likely to benefit from a specific medicinal product is a companion diagnostic: an in vitro diagnostic under the IVDR, with its own classification treatment and a consultation with a medicines authority. It sits on the border between three frameworks — devices, diagnostics and medicines — and is not a case to resolve from a guide.

Systems and procedure packs

Putting an MDR device and an IVD in the same box does not merge the regulations. Each component keeps its own regime, and the person assembling the pack takes on obligations of their own.

Products that are neither

Before choosing between the two regulations, confirm that the product is a medical device at all. Lifestyle and wellness applications, general laboratory equipment not intended for diagnostic use, and research-use-only products sit outside both. Getting this wrong in the optimistic direction — assuming you are outside — is the more expensive error, because it means no framework was applied at all.

A workable order of questions

  • Is it a medical device? If the intended purpose is not medical, neither regulation applies. This is qualification, and it comes before everything else.
  • Does it examine a specimen in vitro? Yes means IVDR, no means MDR. For software, ask what data it acts on.
  • Does it have separable functions? Qualify each one independently before classifying anything.
  • Only then, classify — under the rules of the regulation you have landed in, and against all of them, because the strictest applicable rule prevails in both.

NOTE —

Where the answer is genuinely unclear, the MDCG borderline and classification manual records how comparable cases have been treated, and a Competent Authority can be asked directly. Both are slower than guessing and considerably cheaper.

Common questions

What is the difference between the MDR and the IVDR?

The MDR, Regulation (EU) 2017/745, covers medical devices. The IVDR, Regulation (EU) 2017/746, covers in vitro diagnostic medical devices — those that examine specimens taken from the human body. They were adopted on the same day and share much of their structure, but they have different classification rules and different class labels: I, IIa, IIb and III under the MDR; A, B, C and D under the IVDR.

How do I know if my product is an in vitro diagnostic?

Ask what it examines. If the device works on a specimen taken from the body — blood, urine, tissue, saliva — outside the body, and its purpose is to provide information about a physiological or pathological state, it falls under the IVDR. If it acts on the patient rather than on a specimen, it falls under the MDR.

Which regulation applies to diagnostic software?

It follows the data. Software that interprets the result of an in vitro examination of a specimen is governed by the IVDR. Software that works on other clinical data — images, vital signs, patient records — is governed by the MDR. MDCG 2019-11 is the endorsed guidance on qualifying and classifying software under both regulations.

Are IVDR class letters equivalent to MDR classes?

No. IVDR classes A to D and MDR classes I to III are separate scales built on different criteria. IVDR classification weighs the risk to the individual patient and the risk to public health, which has no equivalent in the MDR. Mapping one onto the other by position produces confident wrong answers.

Does MDClassify cover the IVDR?

Not yet. The engine implements Annex VIII of the MDR only. In vitro diagnostics need their own engine because the IVDR rules turn on what a test detects and the consequence of a wrong result, not on invasiveness or duration of contact. It is on the roadmap.

IF THE ANSWER IS THE MDR

The questionnaire applies the Annex VIII rules and returns the class with every rule that applied and the reasoning behind it. In vitro diagnostics are not covered — that engine is on the roadmap, and we would rather say so than return a number computed under the wrong framework.

Classify under the MDRDo I need a Notified Body?All guides

This guide describes Regulations (EU) 2017/745 and (EU) 2017/746 and is intended for orientation. It is not legal or regulatory advice. Qualification and classification of a specific product should be confirmed by a qualified regulatory affairs professional against the full text of the applicable regulation and the MDCG guidance.