MDClassifyStart free classification

GUIDE

Do I need a Notified Body?

For most devices the answer follows mechanically from the class. The exceptions are where it gets expensive: three sub-categories of Class I pull a Notified Body back in for one narrow aspect, and being self-declared has never meant being unregulated.

Written with Francesca Di Giuseppe, Senior QA/RA Specialist (LinkedIn) — content reviewed on 1 August 2026

The short answer

Whether a Notified Body is involved follows from the class, which is why the class has to be settled first. If you do not have one yet, the questionnaire takes about five minutes.

CLASSNOTIFIED BODYWHAT THEY ASSESS
INot requiredNothing — the manufacturer self-declares
Is · Im · IrRequired, narrowlySterility, the measuring function, or reprocessing — that aspect only, not the whole device
IIaRequiredQuality management system, plus technical documentation for at least one representative device per category
IIbRequiredAs above, per generic device group — and for implantables, every device, subject to the exceptions in Article 52(4)
IIIRequiredFull quality management system and the technical documentation of every device

The Class I exceptions that catch people out

“Class I, so no Notified Body” is right most of the time and expensive when it is wrong. Three sub-categories pull one back in:

  • Class Is — the device is supplied sterile. The Notified Body assesses the establishment, securing and maintenance of sterile conditions.
  • Class Im — the device has a measuring function. The assessment covers the metrological requirements, not the clinical performance.
  • Class Ir — the device is a reusable surgical instrument. The assessment covers reuse: cleaning, disinfection, sterilisation, maintenance and functional testing, and the related instructions.

NOTE —

Class Ir is the newest of the three and the one most often missed, because reusable surgical instruments required no Notified Body at all under the previous directives. Great Britain, which never adopted the MDR, still has no equivalent requirement — see classification by market.

A device can be more than one

A sterile, reusable instrument with a measuring function is Is, Im and Ir at once. The Notified Body's scope is the union of those aspects — still not the whole device, but no longer a narrow review either.

Self-declared does not mean unregulated

This is the most consequential misunderstanding in the whole area. Not needing a Notified Body removes an auditor, not the obligations. For an ordinary Class I device you must still:

  • compile and maintain the technical documentation in Annexes II and III, including the classification rationale;
  • carry out and keep current a clinical evaluation;
  • operate a quality management system proportionate to the device;
  • appoint a person responsible for regulatory compliance under Article 15 — micro and small enterprises may use a permanently available external person rather than an employee;
  • assign a UDI and register the device and the economic operators in Eudamed;
  • run post-market surveillance under Article 84 and produce a PMS report;
  • draw up the EU declaration of conformity and affix the CE marking.

NOTE —

Because no Notified Body reviews the reasoning, an incorrect Class I determination often survives until an authority inspection or an incident — at which point the declaration of conformity, and the CE marking that rests on it, were never valid.

Choosing a Notified Body

Designation scope comes first

Notified Bodies are designated for specific ranges of device types, and may only assess devices inside their designation. The public database of designated bodies lists each one's scope. Check it before making contact: approaching a body that cannot lawfully assess your device is a common way to lose several weeks.

Capacity is the real constraint

There are far fewer designated bodies under the Regulation than there were under the directives, and demand has outstripped them. Waiting lists are a normal part of planning rather than an exception, and a body may decline a client whose portfolio does not fit its priorities.

The practical consequence for a project plan: contact bodies early, expect the conversation to start with a scoping questionnaire rather than a quotation, and treat the contract date rather than the application date as the start of the timeline.

What escalates the scrutiny

  • Implantable Class IIb — technical documentation is assessed for every device, except for the items listed in Article 52(4) such as sutures, staples, dental fillings, screws, wedges, plates, wires, pins, clips and connectors.
  • Class III implantable — the clinical evaluation consultation procedure applies, in which an independent expert panel may review the Notified Body's assessment. It adds materially to the timeline and cannot be compressed.
  • Devices with a medicinal substance — the Notified Body must consult a medicines authority on the substance. Two regulators, two timelines.

If the answer is unwelcome

The instinct on discovering that a device is IIa rather than I is to look for a way back down. There is a legitimate one and an illegitimate one, and they look similar from the inside.

Legitimate: the intended purpose was drafted more broadly than the product warrants. Narrowing it to what the device actually does — and removing the claims that were aspirational — can change the class, because the class follows the intended purpose. This is a real exercise and it is done in the technical documentation, not in the marketing.

Illegitimate: writing an intended purpose that contradicts how the device is sold. A Notified Body reads the website alongside the file, and an authority reads both after an incident.

NOTE —

If you narrow the intended purpose, re-run the classification against the new wording rather than assuming the outcome. The account area keeps both versions and shows exactly which answers changed and whether the class moved.

Common questions

Which medical devices need a Notified Body under the EU MDR?

Every class except Class I. Class IIa, IIb and III all require a Notified Body. Class I devices are self-declared, unless they are supplied sterile, have a measuring function, or are reusable surgical instruments — in which case a Notified Body assesses that specific aspect only.

Can I CE mark a Class I device myself?

Yes. For an ordinary Class I device the manufacturer compiles the technical documentation, performs the clinical evaluation, draws up the EU declaration of conformity and affixes the CE marking without any Notified Body. That is self-declaration, not self-regulation: every other obligation still applies.

What are Class Is, Im and Ir?

Sub-categories of Class I: Is for devices supplied sterile, Im for devices with a measuring function, and Ir for reusable surgical instruments. Each requires a Notified Body, but only for the aspect concerned — sterility, metrological performance, or reprocessing — not for the whole device.

Does a Notified Body decide my device's class?

No. The manufacturer classifies the device and documents the rationale; the Notified Body reviews that rationale as part of conformity assessment and can challenge it. Discovering at that point that the class was wrong is the expensive way to find out.

Can any Notified Body assess any device?

No. Notified Bodies are designated for specific ranges of device types, and can only assess devices within their designation. Approaching one that is not designated for your device type wastes weeks, so check the scope before making contact.

START WITH THE CLASS

The questionnaire returns the class, whether a Notified Body is required, the conformity assessment routes open to you and the obligations attached — in a report you can hand to a regulatory professional. Free, about five minutes.

Classify a deviceRule 11: software classificationAll guides

This guide describes Regulation (EU) 2017/745 and is intended for orientation. It is not legal or regulatory advice. The conformity assessment route for a specific device should be confirmed by a qualified regulatory affairs professional against the full text of the Regulation.